Helmholtz Control System
A complete control system offering active compensation and orthogonality correction for Helmholtz Coils.
Technical specifications
- Field strength
- ±1mT from DC to 440 Hz
- DC disturbance field attenuation
- -46 dB ±1 dB
- Zero field DC drift over 6 hours
- <100nT
- System noise
- <3nT peak to peak
- Bandwidth: HC1, HC2, HC16, & BH1300HF-4-3-A
- DC to 12 Hz -3 dB, one pole LPF
- Bandwidth: BH1300-C, BHC2000-A
- DC to 4 Hz -3 dB, one pole LPF
- Reference magnetometer range
- ±500μT
- Reference magnetometer scaling
- 20 mV/μT (±10 V full scale)
- Maximum field per axis
- <±500μT (applied and disturbing field combined)
- Warm-up time
- 30 min
- Orthogonality error adjustment range
- ±1˚ nominal
- Gain adjustment range
- ±10% nominal
- Offset adjustment range
- ±800nT nominal
- IP Rating
- IP20
About
This system is the perfect companion to Bartington Instruments’ Helmholtz Coils (HC1, HC2 and HC16). It includes the PA1 Power Amplifier, CU1 Control Unit, and CU2 Closed-Loop Module. LabView™-based control software is also supplied to provide a turnkey Helmholtz Coil System. Key features include a field strength of ±1mT from DC to 440Hz, active compensation using the CU2, and orthogonality correction using the PA1. Applications for this system include directional tool calibration, reference magnetometry, compass and magnetometer calibration, and biomagnetics.
Documentation
- DS2613.pdf PDF source: bartingtondownloads.com ↗
- OM3224.pdf PDF source: www.bartingtondownloads.com ↗
- OM3999.pdf PDF source: www.bartingtondownloads.com ↗
Need the full ICD, test reports or a specific revision? Ask the supplier directly.