CryoMag®
CryoMag three-axis fluxgate magnetometers are designed for operation at temperatures as low as 2K, making them ideal for use in cryostats for superconducting radio frequency (SRF) cavities, quantum computing, and other ultra-low temperature environments.
Typed specifications44 values
Discloses 7 of 11 fields engineers filter by in this category (category median 8).
- Mass
- 320 g typical · datasheet p.5 320 g typ.
- Measurement range
- 70 µT abs · datasheet p.4 100 µT CryoMag 100 | 250 µT CryoMag 250 | 500 µT CryoMag 500 | 70 µT CryoMag 70 +3
- Envelope (L×W×H)
- 25.4mm Diameter x 220mm Long Electronics Enclosure · datasheet p.5 7 x 5 x 38mm typical IE probe | 7 x 5 x 38mm typical Probe Dimensions IE | 21mm Diameter x 50mm Long Probe Dimensions Standard | 21mm Diameter x 50mm Long standard probe
- Noise density
- 0.02 nT/√Hz max · at 1Hz · datasheet p.4
- Supply voltage
- 12 V min · datasheet p.5 17 V max | 12–17 V ±12 to 17V
- Operating temperature
- 0 °C min · Electronics + interconnection block (dynamic cable) · datasheet p.4 -40 °C min · Electronics + interconnection block (static cable) | -271 °C min · Probe + probe harness | 70 °C max · Electronics + interconnection block (dynamic cable) | 70 °C max · Electronics + interconnection block (static cable) +4
- Sensor type
- fluxgate datasheet p.2
- Axes
- 3 datasheet p.4
- Output type
- analog ±10V single-ended · datasheet p.4
- Connector
- Fischer AL-1731-DEU-1031-A010-SR-11-11-G-12 datasheet p.5
- Survival temperature
- -40 °C min · datasheet p.4 60 °C max | -40–60 °C storage
- Intended use
- ground or laboratory datasheet p.3
- Configuration
- separate sensor head two part: probe + electronics enclosure · datasheet p.5
- Zero offset
- 30 nT ±100µT range · datasheet p.4 40 nT ±250µT range | 50 nT ±500µT range | 30 nT ±70µT range
- Linearity error
- 0.002 % least squares fit · datasheet p.4
- Axis orthogonality error
- 0.1 ° max · CryoMag (standard 3-axis probe) · datasheet p.4
- Bandwidth
- 2 000 Hz min · -3dB · datasheet p.4
Each value is typed with its unit and measurement condition and cites the datasheet page it was taken from. Nothing is guessed: a field that is absent here has not been published by the manufacturer. machine-extracted ✓ checked by NewSpace Market ✓ manufacturer confirmed by the manufacturer.
As published by the supplier22 lines, in the manufacturer's own wording
- Operating temperature
- down to 2K
- Noise level
- <20pTrms/√Hz at 1 Hz
- Frequency response
- DC to 1 kHz
- Measuring ranges
- ±70, ±100, ±250 and ±500μT
- Number of axes
- Three
- Analogue outputs
- ±10 V single-ended
- Full Scale Measuring Ranges
- ±70µT, ±100µT, ±250µT, ±500µT
- Scaling calibration error
- <0.5%
- Linearity error
- 0.0015% (least squares fit)
- Bandwidth (-3dB)
- >2 kHz (-11 dB/octave roll-off)
- Perming (magnetisation hysteresis)
- <2nT at 1x Full-scale, when powered
- Orthogonality error between axes
- <±0.1°
- Excitation breakthrough
- <5 mV pk-pk at 15.625 kHz typical
- Start-up time
- <100 ms
- Warm-up time
- 15 minutes to meet specifications for scaling
- Storage temperature range
- -40 to +60°C
- Vacuum
- 10e-6 mBar
- IP Rating
- IP40
- Humidity - electronics
- Up to 90% RH, non-condensing
- Probe Dimensions
- 21 mm Diameter x 50 mm Long
- Dimensions – Electronics Enclosure
- 25.4 mm Diameter x 220 mm Long
- Weight
- 320 g typical
About
CryoMag three-axis fluxgate magnetometers are perfect for operation at temperatures as low as 2K, making them the ideal choice for use in cryostats for superconducting radio frequency (SRF) cavities, quantum computing, and other ultra-low temperature environments. These fluxgate magnetometers are available with a standard single probe or multiple sensor heads, all offering high precision measurement of static and alternating magnetic fields. Key features include an operating temperature down to 2K, noise level down to <20pTrms/√Hz at 1Hz, and a frequency response from DC to 1kHz. They offer measuring ranges of ±70, ±100, ±250 and ±500μT, and are available as a 3-axis probe or with 3 single-axis sensor heads (CryoMag-IE). Applications include magnetic field monitoring in cryostats and other low temperature environments, as well as use as feedback sensors in active magnetic field cancellation systems.
Documentation
- DS4476.pdf PDF source: www.bartingtondownloads.com ↗
- OM4476.pdf PDF source: www.bartingtondownloads.com ↗
- DR4476.pdf PDF source: www.bartingtondownloads.com ↗
Need the full ICD, test reports or a specific revision? Ask the supplier directly.