Mag-03RCU
These three-axis range fluxgate magnetometers are designed for magnetic signature measurements and marine surveillance, operating over cables up to 2000m and submersible to 100m.
Typed specifications33 values
Discloses 7 of 11 fields engineers filter by in this category (category median 8).
- Mass
- 1 100 g sensor only · datasheet p.5 1 650 g with bowspring assembly
- Measurement range
- 100 µT datasheet p.4 100 µT ±100μT option | 200 µT ±200μT option | 300 µT ±300μT option | 200 µT +1
- Envelope (L×W×H)
- 582.5 x ø50mm datasheet p.5 582.5 x ø50mm Mag-03RCU | 516 x ø50mm Mag-03RCUT
- Noise density
- 0.02 nT/√Hz max · measurement noise floor at 1Hz · datasheet p.4
- Supply voltage
- 15 V min · Supply voltage · datasheet p.5 12 V min · Supply voltage (absolute min. and max.) | 30 V max · Supply voltage | 30 V max · Supply voltage (absolute min. and max.) | 12–30 V absolute min. and max. +1
- Operating temperature
- -10 °C min · datasheet p.5 50 °C max | -10–50 °C
- Sensor type
- fluxgate datasheet p.2
- Axes
- 3 datasheet p.4
- Output type
- analog ±10V balanced differential · datasheet p.5
- Connector
- Hydrobond HRS106-10P/SS/L, or Hydrobond HRS106-10P/Ti/L magnetometer connector · datasheet p.5
- Intended use
- ground or laboratory magnetic signature measurements in marine ranges, degaussing … · datasheet p.3
- Configuration
- integrated (sensor + electronics) single cylindrical housing · datasheet p.5
- Zero offset
- 30 nT ±100μT range · datasheet p.4 60 nT ±200μT range | 90 nT ±300μT range
- Linearity error
- 0.002 % max · datasheet p.4
- Axis orthogonality error
- 0.2 ° max · all axes to the reference face · datasheet p.4
- Bandwidth
- 3 000 Hz at -3dB (±10%) · datasheet p.4
Each value is typed with its unit and measurement condition and cites the datasheet page it was taken from. Nothing is guessed: a field that is absent here has not been published by the manufacturer. machine-extracted ✓ checked by NewSpace Market ✓ manufacturer confirmed by the manufacturer.
As published by the supplier21 lines, in the manufacturer's own wording
- Measuring range options
- ±100μT, ±200μT or ±300μT
- Bandwidth
- DC to 3 kHz
- Cable lengths
- up to 2 km
- Submersible depth
- 100m
- Number of axes
- Three
- Measuring range
- ±100μT
- Measurement noise floor
- <20pTrms/√Hz at 1 Hz
- Scaling
- 100 mV/μT
- Scaling temperature coefficient
- 10ppm/°C
- Offset
- ±30nT
- Offset temperature coefficient
- ±0.2nT/°C
- Scaling error
- ±0.5% at DC
- Orthogonality error
- <0.2°
- Linearity error
- <0.0015%
- Frequency response
- DC to 1 kHz (±5%) at 1 kHz
- Hysteresis (when powered) within measuring range
- <1nT
- Ingress Protection rating
- IP68 submersible to 100m
- MTBF
- 75000 hours
- Dimensions
- 582.5 x ø50 mm
- Weight
- 1.1 kg (sensor only)
- Supply voltage
- +15 V to +30 V
About
The Mag-03RCU fluxgate magnetometers are engineered for precise magnetic signature measurements and marine surveillance applications. They are capable of operating with cable lengths up to 2000m and are submersible to a depth of 100m. Ease of installation and maintenance is ensured through the availability of underwater mateable connectors, which can be specified in either stainless steel or titanium. These connectors can be integrated directly into the housing or provided on a cable tail. The magnetometers offer measuring range options of ±100μT, ±200μT, or ±300μT, with a bandwidth extending from DC to 3kHz. A built-in test facility is included for convenience. Bartington Instruments’ DAS1 data acquisition system is compatible and can manage up to 80 Mag-03RCU magnetometers simultaneously. Key applications include magnetic signature measurements in marine ranges, degaussing ranges, and general marine surveillance.
Documentation
- DS2276.pdf PDF source: www.bartingtondownloads.com ↗
- OM1004.pdf PDF source: www.bartingtondownloads.com ↗
- DR4144.pdf PDF source: www.bartingtondownloads.com ↗
Need the full ICD, test reports or a specific revision? Ask the supplier directly.