Gradiometer (MFG-2S)
Dual-sensor fluxgate gradiometer for detecting small-scale magnetic field anomalies and spacecraft magnetic moment testing.
Typed specifications31 values
Discloses 10 of 11 fields engineers filter by in this category (category median 8).
- Mass
- 395 g datasheet p.2 395 g box, MFG-2S | 105 g sensor
- Measurement range
- 65 µT datasheet p.2 -65–65 µT can be adjusted
- Envelope (L×W×H)
- 170 x 105 x 33 datasheet p.2 170 x 105 x 33 mm box, MFG-2S | H: 40mm, Cover: 50mm, Socket: 67mm sensor
- Resolution
- 0 nT datasheet p.2
- Output data rate
- 1 Hz datasheet p.2 1–100 Hz sampling rate selectable 1, 10, 50, 100 Hz | 10 Hz | 50 Hz | 100 Hz
- Noise density
- 0.01 nT/√Hz typical, at 1 Hz · datasheet p.1 0.015 nT/√Hz at 1 Hz | 0.015 nT/√Hz max · at 1 Hz
- Supply voltage
- 4.5–9 V supply voltage option · datasheet p.2 9–18 V supply voltage option | 18–36 V supply voltage option
- Power consumption
- 1.5 W datasheet p.2 1.5 W MFG-2S
- Data interface
- RS422 datasheet p.2 RS-422 615384 Baud
- Operating temperature
- -100–200 °C sensor usable temperature range · datasheet p.1
- Sensor type
- fluxgate vector compensated ringcore fluxgate sensor · datasheet p.1
- Axes
- 6 magnetic field components · datasheet p.2 3 tri-axial, X Y Z
- Output type
- digital control and data output via serial interface · datasheet p.2
- Intended use
- other mobile applications with weight and space limitations; drones… · datasheet p.1
- Configuration
- separate sensor head dual sensor configuration, cable length up to 20 m · datasheet p.1
Each value is typed with its unit and measurement condition and cites the datasheet page it was taken from. Nothing is guessed: a field that is absent here has not been published by the manufacturer. machine-extracted ✓ checked by NewSpace Market ✓ manufacturer confirmed by the manufacturer.
As published by the supplier15 lines, in the manufacturer's own wording
- Sensor configuration
- Two independent fluxgate sensors in spatial gradient arrangement
- Measurement type
- Magnetic field gradient / anomaly detection
- Primary use
- Spacecraft magnetic moment determination, magnetic cleanliness testing
- Known users
- ESTEC (ESA), Thales Alenia Space
- Base sensor noise
- < 15 pT/sqrt(Hz) @ 1 Hz (per Magson fluxgate sensor design)
- Box Dimensions
- 170 x 105 x 33 mm
- Box Weight
- 395 g
- Supply voltage
- 4.5-9, 9-18 or 18-36 V
- Power
- 1.5 W
- Range
- ± 65 000 nT
- Resolution
- 7.7486 pT
- Interface
- RS422
- Data rate
- 615384 Baud
- Sampling Rate
- Selectable 1 Hz, 10 Hz, 50 Hz, 100 Hz
- Logging Memory
- 0.5 or 1 GByte Flash
About
The Magson Gradiometer (MFG-2S) measures the magnetic field gradient using a spatial arrangement of two independent Magson fluxgate sensors. This dual-sensor setup is specifically used to detect small-scale anomalies in the ambient magnetic field by comparing readings between the two spatially separated sensors, canceling out uniform background field contributions while isolating localized gradients. The gradiometer is used for magnetic moment determination and magnetic cleanliness testing of spacecraft components at facilities including ESTEC (the European Space Agency’s technology center) and Thales Alenia Space, making it a valuable tool for satellite AIT (assembly, integration, and test) magnetic compatibility campaigns as well as terrestrial geophysical anomaly detection surveys.
Documentation
- MFG-2S_short_data_sheet.pdf PDF source: www.magson.de ↗
- MFG-2S-LC_short_data_sheet.pdf PDF source: www.magson.de ↗
Need the full ICD, test reports or a specific revision? Ask the supplier directly.