Mag-13®
High performance, low noise fluxgate magnetometers for accurate measurements of magnetic fields in various applications, including space.
Typed specifications75 values
Discloses 7 of 11 fields engineers filter by in this category (category median 8).
- Mass
- 82.5 g typ. · Mag-13MC · datasheet p.7 82.5 g Mag-13MC | 950 g Mag-13MCD / MCDFL / MCDT | 222.6 g Mag-13MS | 300 g Mag-13MSS +9
- Measurement range
- 70 µT nom. · datasheet p.5 1 000 µT ±1000µT option | 100 µT ±100µT option | 250 µT ±250µT option | 500 µT ±500µT option +5
- Envelope (L×W×H)
- Ø 25.4 x 203mm Mag-13MC · datasheet p.7 Ø 60 x 301mm Mag-13MCD | Ø60 x 264mm Mag-13MCDFL | Ø60 x 264mm Mag-13MCDT | 32 x 32 x 203mm Mag-13MS +3
- Noise density
- 0.006 nT/√Hz max · low noise (L), @1Hz, 70 and 100µT range only · datasheet p.5 0.01 nT/√Hz max · standard noise, @1Hz | 0.005 nT/√Hz max · very low noise (Q), at 1Hz, 70 and 100μT range only
- Supply voltage
- 12 V min · datasheet p.6 17 V max | 12–17 V ±12 to 17V
- Operating temperature
- -40 °C min · datasheet p.6 70 °C max | -40–70 °C
- Design lifetime
- 6.8 years product page
- Sensor type
- fluxgate datasheet p.2
- Axes
- 3 datasheet p.5
- Output type
- analog ±10V, output impedance 10Ω · datasheet p.6
- Connector
- Fischer AL-1731-DEU-1031-A010-SR-11-11-G-12 Mag-13MC · datasheet p.7 Fischer AL-1731-DEU-1031-A010-SR-11-11-G-12 Mag-13MC, Mag-13MS | SubConn MCBH10F (underwater mateable) Mag-13MCD | MacArtney Type 4622C Cable Mag-13MCDFL | MacArtney Type MCOM10F Mag-13MCDT +5
- Survival temperature
- -40 °C min · datasheet p.6 70 °C max | -40–70 °C storage
- Intended use
- ground or laboratory defence, physics, geophysics, bioelectromagnetics, mineral ex… · datasheet p.3
- Configuration
- separate sensor head Mag-13U-TPU, unpackaged 2-part (probe + electronics, harness … · datasheet p.7 integrated (sensor + electronics) packaged versions MC / MS / MSS / MCD / MCDFL / MCDT / U
- Zero offset
- 50 nT max · ±1000µT range · datasheet p.5 40 nT max · ±500µT range | 30 nT max · ±70 / ±100 / ±250µT ranges
- Linearity error
- 0.002 % least squares fit · datasheet p.5
- Axis orthogonality error
- 0.1 ° max · all packages except MCD / MCDFL · datasheet p.5 0.2 ° max · MCD and MCDFL
- Bandwidth
- 300 Hz max · ±1000µT range, at full scale · datasheet p.5 3 100 Hz max · ±100µT range, at 50µT peak | 2 500 Hz max · ±100µT range, at full scale | 3 000 Hz max · ±250 / ±500 / ±1000µT ranges, at 50µT peak | 1 100 Hz max · ±250µT range, at full scale +3
Each value is typed with its unit and measurement condition and cites the datasheet page it was taken from. Nothing is guessed: a field that is absent here has not been published by the manufacturer. machine-extracted ✓ checked by NewSpace Market ✓ manufacturer confirmed by the manufacturer.
As published by the supplier16 lines, in the manufacturer's own wording
- Number of axes
- Three
- Full Scale Measuring Ranges
- ±70µT ±100µT ±250µT ±500µT ±1000µT
- Scaling error
- ±0.5%
- Linearity error
- 0.0015%
- Frequency response at 50µT peak
- DC to 1 kHz (±5%)
- Noise: Standard
- ≤10pT rms / √Hz @1 Hz
- Zero Field Offset
- ≤±30nT ≤±40nT ≤±50nT
- Orthogonality error between axes
- <±0.1°
- Supply Voltage
- ±12 to 17 V
- Current Consumption – Positive
- 35 to 41 mA
- Current Consumption – Negative
- 13 to 17 mA
- Analogue Output
- ±10 V
- Output Impedance
- 10Ω
- Operating temperature
- -40 to +70°C
- Storage temperature
- -40 to +70°C
- Lifetime
- 60,000 hours
About
Mag-13 three-axis fluxgate magnetometers offer accurate measurements of magnetic fields. They have many applications, from geophysics and mineral exploration, to physics, defence, and space. Mag-13 fluxgate magnetometers are available in multiple enclosure types, and unpackaged, making them suitable for use in the laboratory, in the field, and underwater. They are easy to integrate into your own measurement platform, and can be combined with Bartington’s Power Supplies and Data Acquisition Units to create a complete measurement system. Key features include packaged, unpackaged and submersible models, noise levels down to <5pTrms/√Hz at 1Hz, measuring ranges from ±70 to ±1000µT, and bandwidth to 3kHz.
Documentation
- DS3143.pdf PDF source: www.bartingtondownloads.com ↗
- OM3143.pdf PDF source: www.bartingtondownloads.com ↗
- DR3357.pdf PDF source: www.bartingtondownloads.com ↗
Need the full ICD, test reports or a specific revision? Ask the supplier directly.