Magmeter-2
The Magmeter-2 is a battery-powered unit designed for rapid monitoring of magnetic fields, providing filtered readings of a magnetometer’s X, Y, and Z outputs on an LCD display panel.
Typed specifications30 values
Discloses 6 of 11 fields engineers filter by in this category (category median 8).
- Mass
- 630 g battery included · datasheet p.8
- Measurement range
- 1 000 µT up to ±1000µT · datasheet p.8 100 µT up to ±100µT
- Envelope (L×W×H)
- 106 x 63 x 148mm datasheet p.8
- Resolution
- 100 nT display, for ranges up to ±1000µT · datasheet p.7 10 nT display, for ranges up to ±100µT | 100 nT for ranges up to ±1000µT | 10 nT for ranges up to ±100µT
- Operating temperature
- 0 °C min · datasheet p.8 50 °C max | 40 °C max · for charging | 0–50 °C 0°C to +40°C for charging
- Axes
- 3 datasheet p.7
- Output type
- analog ±10V analogue output, 3 off BNC (X, Y, Z) · datasheet p.7
- Connector
- Hirose RM15TRD-10P sensor connector (mating RM15TPD-10S); 3 off BNC outputs; 2.1mm battery charger inlet datasheet p.7 2.1mm socket Battery charger inlet | 3 off BNC connectors (X, Y and Z) Output connectors | Hirose RM15TRD-10P (mating connector – RM15TPD-10S) Sensor connector
- Survival temperature
- 5 °C min · long term · datasheet p.8 -10 °C min · short term | -20 °C min · without batteries | 30 °C max · long term | 50 °C max · short term +4
- Intended use
- ground or laboratory rapid measurements of magnetic field strength, fieldwork · datasheet p.6
- Linearity error
- 1 % datasheet p.7
- Bandwidth
- 9 500 Hz -3dB, DC to 9.5kHz (DC coupling) · datasheet p.7
Each value is typed with its unit and measurement condition and cites the datasheet page it was taken from. Nothing is guessed: a field that is absent here has not been published by the manufacturer. machine-extracted ✓ checked by NewSpace Market ✓ manufacturer confirmed by the manufacturer.
As published by the supplier20 lines, in the manufacturer's own wording
- Resolution (up to ±100μT)
- 10nT
- Resolution (up to ±1000μT)
- 100nT
- Low-pass filtering
- 9.5 kHz
- High-pass filter
- 0.1 Hz (selectable)
- Number of input channels
- Three
- Voltage input range
- ±10 V
- Bandwidth (-3dB)
- DC to 9.5 kHz (DC coupling); 0.1 Hz to 9.5 kHz (AC coupling)
- Frequency response (±2%)
- DC to 6 kHz (DC coupling) ±2%
- Measurement noise floor
- <200nV/√Hz at 1 Hz (battery powered)
- Linearity
- 1%
- Offset error
- 0.5% full scale
- Operating temperature
- 0°C to +50°C (0°C to +40°C for charging)
- Storage temperature
- +5°C to +30°C long term
- Humidity
- 0–90% non-condensing
- Dimensions
- 106 x 63 x 148 mm
- Weight
- 630 g (battery included)
- Analogue output voltage
- ±10 V
- Battery life
- 8 hours (typical)
- Battery charging time
- <3 hours for full charge
- Output Impedance
- 50Ω per channel
About
The Magmeter-2 enables rapid monitoring of magnetic fields and displays filtered readings from a magnetometer’s X, Y, and Z outputs on an LCD panel. It also offers analogue outputs for each axis, allowing for display or precise data acquisition using third-party equipment. This unit is battery-powered, making it suitable for fieldwork and reducing internal noise for precision measurements. It applies low-pass filtering (9.5kHz) to all inputs to prevent the magnetometer’s excitation signal from affecting measurements. Additionally, a high-pass filter (0.1Hz) is selectable via a front panel switch for AC measurements. The Magmeter-2 is supplied with a mains AC charging adaptor for continuous operation.
Documentation
- DS2520.pdf PDF source: www.bartingtondownloads.com ↗
- OM2448.pdf PDF source: www.bartingtondownloads.com ↗
- DR3694.pdf PDF source: www.bartingtondownloads.com ↗
Need the full ICD, test reports or a specific revision? Ask the supplier directly.